Back to Blog
Metrology Minute

Why Is Non-Contact Inspection Important for Medical Devices?

Fong MH
17 November 2026 2 min read
Engineers viewing a scanned medical component in ZEISS INSPECT beside a rotation table

Medical device components often have intricate geometries, tight dimensional tolerances and delicate surfaces. Non-contact optical 3D metrology measures accessible surfaces without a probe touching the part. This helps protect the surface from contact marks while giving engineers the data they need to verify critical dimensions and check how closely a component follows its intended shape.

Consider the curved surface of a knee implant component. Its width and thickness can meet specification even when a section of the contour deviates from the design. A full-field 3D scan helps engineers locate that variation, see how far it extends and assess whether it falls within the specified profile tolerance.

What does a full-field 3D scan reveal?

An optical 3D scanner captures the component from multiple angles to build a digital representation of its surface. Using ZEISS INSPECT Optical 3D, engineers can align the scan with the reference CAD model and display the differences as a colour-coded deviation map.

On an implant casting, for example, the map can reveal a raised area and show how it blends into the surrounding contour. An engineer can use those measurements to check the feature against the drawing and, if rework is needed, show the production team exactly where to focus.

From prototype inspection to production monitoring

During development, a scan-to-CAD comparison helps engineers verify how closely a prototype reproduces the intended geometry. Once the component enters production, inspection templates allow quality teams to evaluate the same features consistently across successive parts. Trend analysis can then show whether a dimension or surface profile is changing over time, giving engineers a specific feature to investigate if a deviation recurs.

The stage of manufacture also influences the choice of inspection method. Optical scanning can capture the freeform geometry of an implant casting, while a mirror-like finish after polishing can make optical measurement more challenging. Engineers therefore need to consider the surface finish alongside the feature and its tolerance when selecting a measurement approach.

How Orchid inspects knee implant components

Orchid Orthopedic Solutions manufactures cobalt-chromium knee implant components through investment casting. Their curved shapes made single-point inspection time-consuming and limited the number of features the team could check. After evaluating measurement repeatability and reproducibility, Orchid adopted ZEISS ScanBox with ZEISS INSPECT and worked with ZEISS to automate its inspection routine.

Previously, Orchid needed around four minutes per component to measure four features. With the automated workflow, the team could inspect 40 components in 45 to 60 minutes while measuring more features on each one. The 3D surface maps also helped engineers identify areas requiring rework and explain their findings to customers.

Speak to FIRST about your medical device component

Share the geometry, surface finish and tolerances of your component. Our team can help you select a ZEISS metrology solution suited to your inspection requirements.

Written by
Principal Application Engineer, FIRST Metrology
ZEISS-Certified Metrology Specialist · B.Eng (Hons) Mechanical Engineering

Fong MH is Principal Application Engineer at FIRST Metrology, bringing over 15 years of experience in industrial metrology, optical 3D measurement and quality assurance. As a ZEISS-certified metrology specialist, he has extensive hands-on experience in 3D scanning, automated inspection, GD&T evaluation and ZEISS metrology solutions, helping manufacturers solve complex measurement challenges and achieve accurate, repeatable and efficient quality inspection.